Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT200900025

LightSpeed 3.X (Ultra, Plus, QXi) - GRE

Symptom

Connect Pro Modality Work List will not update. Connect Pro Modality Work List will not update

Error message

Error: 200900025 Generic Patient Worklist Error , See Long Description for full errors See Long Description for errors The error can be in different contexts per the below Thu Jan 4 10:12:21 2007 Error: 200900025 pct01 WLServer WLapiSystem.c 190 Function: Data Management : OC Processing Patient Worklist Manager error. Source File: WLServer.c, Line: 1609, Status: 28 Fatal Unknown DCM ERROR Mon Sep 19 08:28:09 2005 Error: 200900025 ct01 piRxMgmt WLapiSystem.c 190 Function: Data Management : OC Processing Patient Worklist Manager error Source File: WLapiWorklistFile.c, Line: 6038, Status: 3 WLgetTime(): Could not read from /usr/g/db/patientInfo/Update.db. PS000007990.db for reading. Error: 200900025 Host: ls08 Process: WLServer File: WLapiSystem.c Line: 190 Function: Data Management : OC Processing Patient Worklist Manager error. Source File: WLS_mfm.c, Line: 1465, Status: 23 Parse Failed : Patient Worklist Manager HIS(40)(3) :-104 Error: 200900025 Process: WLServer File: WLapiSystem.c Line: 190 Function: Data Management : OC Processing Patient Worklist Manager error. Source File: WLServer.c, Line: 293, Status: 19 DICOM Association Inactivity Time-out Tue Sep 26 09:51:49 2006 Error: 2

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.