Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230023045230017823038304900000230017801

LightSpeed 4.X - Octane

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017823 Error: Invalid ratio range detected during the Fast Cal See the long description for the full error Thu Aug 26 02:56:40 2004 Error: 230017823 Host: ct01 Process: accs File: collSweepProcessing.c Line: 2436 ACCS CCal: Error: Invalid ratio range detected during the Fast Cal. The minimum ratio is 1.038304 and is great than the maximum ratio 0.900000

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405I.3_H4.0M5.1 for LightSpeed 4.x LS2002M5.1 405I.3_H4.0M5.1 for LightSpeed 4.X LS2002M5.1 (IRIX) LS2002M5.1 Supports New Motorola-bd P/N 2370825. FRU Collector 5114901 (Motorola-bd 2370825 + Instructions 5114494-100) will be available around FW30, 2004 GOC1 Console OPERATING SW: Software Collector 5112672 Patch C.1 5113667 CLASS C: 2395172 or 2401759 (6 pack) 1104I.2_H4.0M5CLASSC CLASS M: 2395173 or 2401759 (6 pack) 1204I2_H4.0M5CLASSM 405I.3_H4.0M5.1 for LightSpeed 4.X LS2002M5.1 (IRIX) LS2002M5.1 Supports New Motorola-bd P/N 2370825. FRU Collector 5114901 (Motorola-bd 2370825 + Instructions 5114494-100) will be available around FW30, 2004 GOC1 Console OPERATING SW: Software Collector 5112672 Patch C.1 5113667 CLASS C: 2395172 or 2401759 (6 pack) 1104I.2_H4.0M5CLASSC CLASS M: 2395173 or 2401759 (6 pack) 1204I2_H4.0M5CLASSM 5112672 LIGHTSPEED 16 MDAS M5.1 Collector Number Quantity Description/Name, 5112617 1.0 PC LS2002M5.1 FOR H4 5112618 1.0 PC LS2002M5.1 FOR H4 APPS DISK 5112619 1.0 PC LIGHTSPEED PATCH DISK
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.