GE Healthcare · CTLightSpeedVCT2601151902601151917156429999260100042
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260115190 Cradle encoder fault See the long description for the full error Wed May 25 17:35:00 2005 Error: 260115190 Host: tgp Process: tgp_subsys File: mtcb_cmc_isrfunc.c Line: 163 Wed May 25 17:35:00 2005 Host : Table Sub-System Ermes # : 260115190 Exception Class : Soft Severity : Pri/Soft File : mtcb_cmc_isrfunc.c Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle encoder fault . Error: 260115191 Cradle position is out of tolerance See long description for full error Thu Sep 1 09:04:12 2005 Error: 260115191 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 414 Thu Sep 1 09:04:12 2005 Host : Table Sub-System Ermes # : 260115191 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle position is out of tolerance. cradle_position(um) : 1666 target_position(um) : 71564 velocity (um/s) : 29999 fpga status : 0x8140 Error: 260100042 Cradle was NOT ready See the long description for the full error code Tue May 17 16:02:03 2005 Error: 260100042 Host: tgp Process: tgp_subsys File: scHdwRdy.c Line: 70 Tue May 17 1
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5122207-2 ABSOLUTE ENCODER FOR CRADLE GT (replaces 5122207) See long description for details Part #: 5122207-2 ABSOLUTE ENCODER FOR CRADLE GT (replaces 5122207) Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: ABSOLUTE ENCODER FOR CRADLE GT FRU: Yes Status: Orderable Comments: This part is used in GT-1700mm and GT-2000mm tables. Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required