Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCTBrightSpeed26011821550368

LightSpeed 5.X (60kW)

Symptom

Scanner hardware stopped scan with KV errors. Scanner hardware stopped scan with Error: 260118215 (60-0318)kV did not reach 75 percent after 20ms.

Error message

Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Mar 25 15:35:27 2006 Host : Jedi Ermes # : 260118215 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50368/1/3 (60-0318)kV did not reach 75 percent after 20ms. Indicates that there was no HV ON signal generated for exposure time count-up.

Applies to

Models
LightSpeed, VCT, BrightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues