LightSpeed 4.X - GRE
Symptom
Lost control of trackball on operator console during image viewing. OS Kernel aborts and crashes Linux. See Long Description for more Details Lost control of trackball on operator console during image viewing with Error: kernel: keyboard.c: can't emulate rawmode for keycode 272, 273, or 274, then OS Kernel aborts and crashes Linux. [insite@ct01 config]$ tail -420f /var/log/messages Nov 22 13:07:31 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 273 Nov 22 13:07:36 ct01 last message repeated 59 times Nov 22 13:07:37 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272 Nov 22 13:07:37 ct01 last message repeated 3 times Nov 22 13:07:37 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 274 Nov 22 13:07:41 ct01 last message repeated 7 times [insite@ct01 log]$ egrep "can't|data|error|fail|hard|panic|retry|soft" /var/log/messages* | more /var/log/messages .1:Nov 22 13:55:57 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 273 /var/log/messages.1:Nov 22 13:56:08 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272 Other Issues seen when this error is found in message log: System takes a long time booting up and is sluggish after the boot up.
Error message
kernel: keyboard.c: can't emulate rawmode for keycode 273 insite@ct01 config]$ tail -420f /var/log/messages Nov 22 13:07:31 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 273 kernel: keyboard.c: can't emulate rawmode for keycode 272 insite@ct01 config]$ tail -420f /var/log/messages Nov 22 13:07:31 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
- Area
- CT
Full cause & step-by-step resolution