GE Healthcare · CTLightSpeed182404
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
After hardware reset table elevation moves up/down slightly and stops. See long description for details After hardware reset table elevation moves up/down slightly and stops. No Index Pulse between Interference switches. Table elevation display stays a flashing 888, never references itself.
Error message
Error: 182404 HW error: No Index Pulse between Interference switches. See LD for error as seen in viewlog Error: 182404 HW error: No Index Pulse between Interference switches Wed Aug 13 08:05:33 2003 Error: 182404 Host: ETC Process: Elevation File: emcIntfSwth.c 1.28 10/2/92 Line: 158 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown 030813084201551 Exception Level: Pri/Soft Time: 08:05:33:575 Log Series: 36 HW error: No Index Pulse between Interference switches.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269173 ELEVATION ENCODER ASM FOR H2 LCC -Zeus Table see Long Description for details and picture Part #: 2269173 Part Type: GP CONSUMABLES Part Weight: 3.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER ASM FOR H2 LCC FRU: Yes Status: Orderable Comments: FLS - ELEVATION ENCODER ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296633G3).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 WILL START DEPLOYMENT FROM ASIA ON FW12, 02. MS30544 1/31/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required