GE Healthcare · CTHiSpeed18310918763
HiSpeed CT/i 4.X with Indigo and HSA Gantry
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Method: No Method Line: 248 Function: XRAY GENERATION : KV Loop Scan: 18763/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 15:38:16:120 Log Series: 991 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
- Area
- CT
Full cause & step-by-step resolution