GE Healthcare · CTHiSpeedLightSpeed1832895000118310710785
Scanner Hardware Stopped Scan during Axial Scan with High Voltage errors.
Symptom
Scanner Hardware Stopped Scan with High Voltage errors
Error message
Error: 183289 Unexpected Loss Of High Voltage Wed Aug 4 20:21:15 2004 Error: 183289 Host: OBC Process: Generator File: gen xray.c 1.31 Line: 528 Function: XRAY GENERATION : Rotating HVON Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:21:16:325 Log Series: 59 Unexpected Loss Of High Voltage. . Error: 183107 KV fault detected: Anode shoot-thru. See long description for complete error Thu Mar 10 12:46:26 2005 Error: 183107 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 239 Function: XRAY GENERATION : KV Loop Scan: 10785/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 12:46:26:065 Log Series: 608 KV fault detected: Anode shoot-thru. Fault latch address = ffcff3 hex, Bit = D2
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Inverter) 2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Invertor) 2122960 COMMON INVERTER HSA / RP-EMC FOR ANODE OR CATHODE (CTi Inverter) Part #: 2122960 Part Type: GP REPAIRABLES Part Weight: 30.86 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: COMMON INVERTER FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution