Pular para o conteúdo principal
GE Healthcare · CTLightSpeedBrightSpeed1829005041218512226000682150431

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182900 Stopping scan because a tableside emergency off button has been pressed Open Long Description to see example error Sun Jul 18 10:02:54 2004 Error: 182900 Host: ETC Process: Table Controls File: tc_scan.c 1.14 1/11/94 Line: 235 Function: No System Function Reported Scan: 50412/1/3 Type: Axial Exception Level: Pri/Most Severe Time: 10:02:54:710 Log Series: 57 Stopping scan because a tableside emergency off button has been pressed

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Error: 260006821 Gantry (Axial) LOOP CONTACTOR state is Still OPEN. The Loop Contactor status is NOT changing to the CLOSED state See Long Description for the full error code. Wed Jul 6 07:25:35 2005 Error: 260006821 Host: STC Process: Axial File: a

Applies to

Models
LightSpeed, BrightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues