Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26012101220606

LightSpeed 5.X (80/100kW)

Symptom

Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then stops

Error message

Error: 260121012 Gantry velocity greater than the Maximum allowed velocity. Gantry velocity:2004 mdeg/sec Current tilt position:-20606 mdeg Maximum allowable velocity:2000 mdeg/sec Error: 260121012 Host: tgp Process: tgp_subsys File: tilt_mgr.c Line: 1143 Sat Apr 30 06:22:02 2005 Host : Table Gantry Processor Ermes # : 260121012 Exception Class : Soft Severity : Pri/Soft File : tilt_mgr.c Line# : 1143 Function : No System Function Reported Scan Type : Scout Scan: 1779/3/1 Gantry velocity greater than the Maximum allowed velocity. Gantry velocity:2004 mdeg/sec Current tilt position:-20606 mdeg Maximum allowable velocity:2000 mdeg/sec

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X see LD for details P/N 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X picture below Part #: 2348298 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TIMING BELT, 2.03 PITCH X 6 WIDE FRU: Yes Status: Orderable Parts Offered as: Current: 2348298 Harvested: Not Available Exchange: Not Available Comments: FLS - TIMING BELT. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19. MS20544 ON 3/24/03
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.