Pular para o conteúdo principal
GE Healthcare · CTLightSpeedBrightSpeedHiSpeed260118212

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 260118212 ( 60-0314 ) ILR Current time out. The current resonant frequency is lower than expected See LD for a sample error. Wed Dec 1 18:17:05 2004 Error: 260118212 Host: orp Process: orp_subsys File: gen_exec.c Line: 2148 Wed Dec 1 18:17:04 2004 Host : Jedi Ermes # : 260118212 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2148 Function : No System Function Reported Scan Type : Helical Scan: 1375/2/1 (60-0314)ILR Current time out. The current resonant frequency is lower than expected.

Applies to

Models
LightSpeed, BrightSpeed, HiSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2401826-2 Jedi 60 Shielded Power Unit / High Voltage Tank Assembly FRU Package for G-DAS systems Includes: 2377668-2 Jedi 60 Power Unit 2186730 Jedi 60 HV Tank Shielding improved for G-DAS systems See LD for details 2401826-2 JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FOR G-DAS SYSTEMS (Backward compatible with M-DAS systems) Part #: 2401826-2 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FRU: Yes Status: Orderable Comments: Revised per to MFG confirmation. This is part of Jedi Collector # 2371333-2 INCLUDES: 2377668-2 JEDI 60DC HLB POWER UNIT with INVERTER and PPC KV Control 2186730 JEDI 60 STD HV TANK
Area
CT

Full cause & step-by-step resolution