Unable to Scan with Multi Coil TR Driver Faults Due to Multi Coil Port Connector in the Carriage

Symptom

Unable to Scan with Multi Coil TR Driver Faults

Error message

2225558 Multi-Coil T/R Driver Fault Multi-Coil Path Short Tue Jun 20 08:41:31 2006 Error: 2225558 Host: sommri Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6310 Multi-Coil T/R Driver Fault, Multi-Coil Path Short A short circuit was detected on port 4. Check cable to multi-coil board or check for a shorted pin diode.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR