Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000650050038

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

Gantry will not rotate or scan, alignment lights will not turn ON

Error message

Error: 260006500 Axial AC Drive reported fault code: F60 - Encoder loss. Encoder Loss indicates that the axial drive can not sense the axial encoder See long description for full error Sun Jan 08 11:00:27 2006 Error: 260006500 Host: STC Process: Axial File: AX2fault.c Line: 196 Function: AXIAL CONTROL : axial control Scan: 50038/8/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:00:26:535 Log Series: 35 The axial AC Drive reported fault code: F60 - Encoder Loss This indicates that the axial drive can not sense the axial encoder. Possible causes are: Encoder, cabling, unplugged axial control board or improper axial drive encoder jumper setting.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2218570 BOX OF FUSES for HSA, CTI, and LS 1X - 4X Systems See long description for details 2218570 BOX OF FUSES for HSA, CTI, and LS 1x - 4x Systems Fuse KIT
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues