GE Healthcare · MR2225113

The Upper Coldhead Does Not Turn Off When Scanning.

Symptom

The Upper Coldhead Does Not Turn Off When Scanning. See long description. Coldhead cycling is not working, the upper coldhead does not turn off when scanning. FE did the setup on the SHEM and had the OLC successfully run a Checkout. Can telnet between the MR and the SHEM both ways. The magnet pressure and HE level are good so this should not be causing the failure to cycle the two stage coldhead. The following error is seen in the /usr/g/service/log/tcuErrors log: 06/10/2004 13:04: PMColdHead.cc:29 Comm with SHeM failed 06/10/2004 13:04: PMHost.cc:425 CH Status unchanged: Previous Status = Communication Fail, Current Status = Communication Fail

Error message

2225113 ColdHead switching not enabled during scan. Tue Dec 23 12:17:25 2003 Error: 2225113 Host: lx-mr Process: NSP File: HPC_PORT_ID:hpc_change_state Line: 40 ColdHead switching not enabled during scan. HPC was told to initiate a scan with the ColdHeads on. Image degradation is likely while this condition remains.

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.