Pular para o conteúdo principal
GE Healthcare · MRSigna2225551

Cannot Scan With TR and Dynamic Disable Faults.

Symptom

Prescan is Failing with T/R Driver Faults on All Ports.

Error message

2225551 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 1. See Long Description Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225551 File: PMH.m 1.29 Method: spi_tr_mon Line: 6230 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 1. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues