Scan Aborts with MDS Link Error

Symptom

Scan Aborts with MDS Link Error

Error message

2248664 - MDS link failure. An MDS device failed to respond to SCP Mon Feb 2 13:27:26 2004 2248664 1 MR-OC0 NSP SCP:MdsLinkHWO 1395 MDS link failure. An MDS device failed to respond to SCP. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0025 Transmitted header: 0x10 Gradient Transmitted opcode: 0x02 EN 335

2248674 MDS link failure The MDS fiber optic link appears to be broken between the RF and System cabinets. Mon Feb 2 13:27:28 2004 2248674 1 MR-OC0 NSP SCP:../MdsLinkHWO.cpp 983 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.-> Mon Feb 2 13:27:28 2004 2248674 1 MR-OC0 NSP SCP:../MdsLinkHWO.cpp 983 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.

2248670 MDS link failure. An MDS device failed to respond to SCP on a second

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues