Pular para o conteúdo principal
GE Healthcare · CTHiSpeed186328

Scanner Hardware Stopped Scan with Servo Amp alarm

Symptom

Scanner Hardware Stopped Scan with Servo Amp alarm Scanner Hardware Stopped Scan with Servo Amp alarm. No further details Scanner Hardware Stopped Scan with Servo Amp alarm

Error message

Error: 186328 13-1002-04 Gantry Req Error: ??? Servo amp Alarm is detected See the Long Description for the complete error. Error: 186328 13-1002-04 Gantry Req Error: ??? Servo Amp alarm is detected Gantry Req Error: 4 or 7 Code: 13-1002-04 Name: Servo Amp Alarm. (GP) Class: Error Description: Servo Amp alarm is detected. Potential Cause / FRU 1. Too many acceleration/deceleration in short duration. 2. Servo Amp 3. Axial Motor

Applies to

Models
HiSpeed
Product
HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
Component
2241899 Axial Servo Amp - used on YMS Hispeed with Performix Tube and CTe See long description for details Part #: 2241899 Part Type: GP CONSUMABLES Part Weight: 7.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9237SB SV AMP FRU: Yes Status: Orderable Comments: FLS - SERVO AMP U/O LXI,FXI,DXI PER PCN CT-979NP 2241899 BACKWARD COMPATIBLE TO 2189124. MS08992 DO NOT DISCOUNT - PER SHAKUN GELS - REASON A CURRENT MOTOR DRIVER (2189124) DOESN'T HAVE ENOUGH MARGIN FOR A REGENERATIVE BREAKING. PROCEDURE,AXIAL SERVO AMP LT1
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.