The PDU is in Stop Mode and the EMO Reset Will Not Take Hold Due to Phoenix PDU control board

Symptom

The PDU is in Stop Mode and the EMO Reset Will Not Work See Long Description Sometimes when you push the green EMO reset button on the PDU nothing happens. You cannot hear the 24VDC relay click on nor hear any fans start up. In addition, you keep getting an error message that the PDU is in stop mode or the 24VDC supply has failed after you reset the TPS and attempt to scan.

Error message

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2228875-2 Phoenix PDU Control Board (new part #2330230) 2330230 REPLACES 2228875-2, PARTS ARE FORWARD BACKWARD COMPATIBLE. 2228875-2 EOPL, replaced by 2330230 which is backward compatable
Area
MR

Full cause & step-by-step resolution