GE Healthcare · CTLightSpeed23001708850019230023033
LightSpeed 3.X (Ultra, Plus, QXi) - Octane
Symptom
Fast Calibration fails with many different errors. Fast Calibration fails with many different errors
Error message
Error: 230017088 The mylar window check failed See Long Description for complete error Mon Aug 23 05:50:31 2004 Error: 230017088 Host: usafaultra Process: accs File: aircal processing.c Line: 1025 The mylar window check failed - Scan 50019/1/3 Error: 230023033 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts. See Long Description for complete error Sat Jan 8 08:06:05 2005 230023033 Host: uhct Process: dailyPrepRx File: DailyPrepUtils.c Line: 2967 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required