GE Healthcare · CTLightSpeed230017124550000
40937 After Fast Cal completes software begins a Z-Slope Calibration Test.
Symptom
System fails fastcals with z-slope errors
Error message
Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. See LD for Full Error Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. Tue Dec 27 14:51:03 2005 Error: 230017124 Host: ct01 Process: accs File: zslope.c Line: 1910 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
- Area
- CT