GE Healthcare · CTLightSpeedVCT260107705
LightSpeed 5.X (60kW)
Symptom
Gantry does not rotate. Customer noticed that gantry does not rotate.
Error message
Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. See LD for complete error Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Mon Mar 14 10:34:57 2005 Error: 260107705 Host: tgp Process: tgp_subsys File: scanbusDriver.c Line: 233 Mon Mar 14 10:34:57 2005 Host : Table Gantry Processor Ermes # : 260107705 Exception Class : Abort Severity : Secondary File : scanbusDriver.c Line# : 233 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Retries used: 0
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570-2 H-POWER FUSE KIT used on Lightspeed 5x systems. See Long Description for contents of fuse kit. Part #: 2218570-2 Part Type: GP CONSUMABLES Part Weight: 43.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER FUSE KIT FRU: Yes Status: Orderable Comments: Not Available Fuse KIT contents Mark Number Name Qty Description 1 2351492 2.0 200A FUSE 2 2364059 2.0 GLASS FUSE 3 46-170021P50 2.0 "FUSE 12 AMPS 250 VOLTS BUSSMAN MDA12 DUAL ELEMENT." 4 46-170021P106 2.0 FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 5 2336517-2 2.0 FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 46-170021P52 3.0 "FUSE 3 AMPS 250 VOLTS(REVIEWED TK, 7/91)." 7 46-170021P10 2.0 "006.000A 0250V 3AG FAST UL/C 8 46-170021P101 1.0 "FUSE 20 AMPS 700 VOLTS 2~ LONG X 9/16~ DIA FAST BLOW" 9 46-170021P11 2.0 "7.0A, 250V SLO-BLO FUSE. *TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 10 46-170021P31 2.0 "1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 11 2106993-5 2.0 FUSE 20 AMPS 500 VOLTS .3 SECONDS 12 46-327160P1 2.0 "12.0A, 125VAC, DUAL ELEMENT TIME DLY FNM-12, 0.406 X 1.5 LG, FIBER BODY" 13 2379651 1.0 FUSE - 700 V, 200 A, FAST SEMI
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required