GE Healthcare · CTLightSpeedVCT268722558816047
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
System posts: Edit Patient Data cannot remove the exam entry: it may be in use. System posts Edit Patient Error message: Edit Patient Data cannot remove the exam entry: it may be in use.
Error message
Error: 1031 stored procedure failed See Long Description for Details. Sun Nov 14 20:00:37 2004 Error: 1031 Host: mcas Process: dbrserver stored procedure failed : -284 - getexamuidinfo failed Also check usr/g/ctuser/logfiles/epdlog with command: tail -100 /usr/g/ctuser/logfiles/epdlog | more MESSAGE from Process 26872>> Sun Nov 14 20:12:15 2004 (netclient.c:93) [Client]> The request manager started MESSAGE from Process 26872>> Sun Nov 14 20:12:48 2004 (:0) [EPD] Failed to remove the exam entry Thu Jul 17 13:45:18 2008 Error: 1031 Host: ct99 Process: dbrserver File: batch.ec Line: 599 stored procedure failed : -243 - Fetch failed for image - ISAM code : -143 Wed Jul 2 10:54:38 2008 Error: 1031 Host: CT01 Process: dbwserver File: batch.ec Line: 934 stored procedure failed : -25588 - Identifyimage failed - ISAM code : 22, object , series $ Error: 16047 [EPD] Failed to remove the exam entry See the LongDescription for the complete error code Error: 16047 [EPD] Failed to remove the exam entry Sat Nov 27 19:30:24 2004 Error: 16047 Host: ct04 Process: editPatientData File: Line: 0 [EPD] Failed to remove the exam entry
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required