Scanning aborted with collimator errors

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003100 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um Mon Dec 6 13:01:26 2004 Error: 260003100 Host: HSC Process: CCB Firmware File: ccb_drive_timers.c Line: 277 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 13:01:27:185 Log Series: 42 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296259P6 Rotating Coupling See Long Description for FEMC and pics 46-296259P6 ROCOM COUPLING_BORES 5 MM pictures Part #: 46-296259P6 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: ROCOM COUPLING_BORES 5 MM, 0.250 IN. FRU: Yes Status: Orderable Comments: FLS - ADAPTER ADDED PER PCN #220365.USED ON GANTRY. PER PCN COMMENTS, FRU WILL BE CHANGED FROM N TO 1. MSA8429 01/15/02 COUPLING OF 2 SHAFTS FOR COLLIMATOR MOTOR MS10147 01/02 FOR ROTATING APTERETURE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.