GE Healthcare · MRSigna2241517
Will Not Scan with Gradient Errors 2241517 and 2151183
Symptom
Will Not Scan with Gradient Errors.
Error message
2241517 The GIP reported a Transistor Fault on the X Axis of the GPM. This Fault Could be Related to Noise. Tue Apr 20 20:24:32 2004 Error: 2241517 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2218 The GIP reported a Transistor Fault on the X Axis of the GPM. This fault could be related to noise. Possible causes: Cable between the GIP and GPM, X Axis GPM Main Board or Output Shelf.->
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2151183 GRADIENT POWER MODULE For SGD. 8651 Is SGD Hi-Slew Gradient Power Module.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.