System stopped scanning in the middle of fastcal; operator was unable to cancel or exit fastcal GUI
Symptom
System stopped scanning in the middle of fastcal; operator was unable to cancel or exit fastcal GUI
Error message
Error: 200300013 The cal_proc process returned a bad status of xxx See LD for Full Error Tue May 30 06:42:09 2006 Error: 200300013 Host: dicct Process: cal_composite File: cal_comp_events.c Line: 297 The cal_proc process returned a bad status of 116 Error: 200306004 Cal processing has run out of retry attempts See LD for Full Error Tue May 30 06:42:09 2006 Error: 200306004 Host: dicct Process: cal_composite File: cal_comp_events.c Line: 356 Cal processing has run out of retry attempts Error: 200306019 cal_comp_resp: Execute Cal/Fast cal with scale smartbeam Command is Error code 200300013 See LD for Full Error Tue May 30 06:42:09 2006 Error: 200306019 Host: dicct Process: calibrationRx File: CalRxFastCalMgr.c Line: 994 cal_comp_resp: Execute Cal/Fast cal with scale smartbeam Command is Error code 200300013
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 4.4.19.1 for HiSpeed CT/i with with CT/I (RHAPSODE) CONSOLE (Indigo) - showprods returns Rhapsode Product 4.4.19.1 Intro. Date 10/18/00 Mandatory FMI 25268 Rhapsode Product 4.4.19.1 software for HiSpeed CT/i with SGI Indigo computer Intro. Date 10/18/00 Mandatory FMI 25268 Application Software Set Set Part No. . 2283723-3 Apps. CD Part No 2282440-3 Apps. Ver. 4.4.19M4 Apps. CD Label HiSpeed CT/I App SW showprods/Service Desktop Rhapsode Product 4.4.19.1 Class C CD Part No. Ver. CD Label 2282441-2 4.419M4 HiSpeed CT/I Class C Software Class M CD Part No. Ver. CD Label 2282442-2 4.419M4 HiSpeed CT/I Class M Software Patch Info: P/N 2318818, 4.5 Patch Rhapsode Product 4.4.19.1 Rhapsode patch 4.5.3 ver 1 11/21/2001 Optional FMI 25287A Also available via AutoSC ,_,Number:,2283723-3_______, ,BOM,Rev:,___,PI Rev:,___,CR Nb/Typ:,_______,___, ,CTZ,US, ,JAROSCH,ANN,K , ,Rev:,000,PI Rev:,000,CR Nb/Typ:, , , ,Description/Name:,E,HISPEED CT/I OPER SFWR VER 4.4.19 M4, ,Date:,2000-11-13, ,AIL:, , ,R,S,S,F, ,0001,/,0002, ,Mark, ,Number, ,Quantity,,U/M,U,T,P,C,Description/Name, ,__,1 ,2282439-2 , 1.000,PC ,S,N,N,N,IRIX 6.2M4 OS FOR 4.4 SOF, ,__,2 ,2282440-3 , 1.000,PC ,S,N,1,1,HISPEED CT/I APP SFWR VER, , , , , , , , , , , ,
- Area
- CT
Full cause & step-by-step resolution