GE Healthcare · CTLightSpeed200108008200002352200108051
LightSpeed 4.X - GRE
Symptom
Applications software fails to boot up completely and automatically shuts itself back down. Applications software fails to boot up completely and automatically shuts itself back down
Error message
Error: 200108008 You will not be able to scan because no calibration information is available See the long description for the full error code Thu Jun 9 08:19:31 2005 Error: 200108008 Host: pmct Process: scanRx File: SIRx.cxx Line: 1307 Function: Data Acquisition: OC Processing Failed to get Calibration status. You will not be able to scan because no calibration information is available. Try pressing End Exam followed by New Patient . This will cause ScanRx to attempt to read the calibration database again Error: 200002352 A HOST_SHUTDOWN was initiated by the death of scanmgr See Long Description for full error Tue Nov 23 17:26:47 2004 Error: 200002352 Host: srct Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiated by the death of scanmgr With 06MW03.X software this is listed as: Fri Oct 6 19:02:31 2006 Error: 200002352 Host: dominct Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanmgr. Error: 200108051 An error had been detected while calling getAnd CheckCals function. see LD for full error Sun Oct 23 08:42:11 2005 Error: 200108051 Host: emc-ct1 Process: scanRx File: SIRx.cxx Line: 265
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.