Pular para o conteúdo principal
GE Healthcare · MRSigna2238685223868622386872224666

Will Not Scan with GAP was Unsucessful accessing Gram on all Axis Due to Digital Tuning Board

Symptom

Will Not Scan with Gradient Errors.

Error message

2238685 GAP was unsuccessful accessing X Axis GRAM Digital Tuning Board Mon May 22 07:42:30 2006 Error: 2238685 Host: NMR1 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 299 GAP was unsuccessful accessing X Axis GRAM Digital Tuning Board.^A Check connections to X Axis GRAM Digital Tuning Board. If there is no GRAM, or all GRAMs have Analog Tuning Boards, change digiTuningBoard to `no' in the MRconfig.cfg file and reset TPS

2238686 GAP was unsuccessful accessing Y Axis GRAM Digital Tuning Board Mon May 22 07:42:30 2006 Error: 2238686 Host: NMR1 Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 299 GAP was unsuccessful accessing Y Axis GRAM Digital Tuning Board.^A Check connections to Y Axis GRAM Digital Tuning Board. If there is no GRAM, or all GRAMs have Analog Tuning Boards, change digiTuningBoard to `no' in the MRconfig.cfg file and reset TPS.

2238687 GAP was unsuccessful accessing Z Axis GRAM Digital Tuning Board GAP was unsuccessful accessing Z Axis GRAM Digital Tuning Board. Check connections to Z Axis GRAM Digital Tuning Board. If there is no GRAM, or all GRAMs have Analog Tuning Boards, change digiTuningBoard to `no' in the MRconfig.cfg file and reset TPS.

2224666 Auto Pre

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.