GE Healthcare · CTLightSpeed18260751200
LightSpeed 5.X (80/100kW)
Symptom
Gantry not tilting in either direction
Error message
Error: 182607 HW/SW error: Tilt Velocity tracking See LD for more details Thu May 11 23:04:54 2006 Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 51200/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 23:04:53:540 Log Series: 144 HW/SW error: Velocity tracking error. Actual Velocity 150 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 40 (.01 deg/sec) Cur Pos -1194 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD) . INFO: Probable Problem could be 180V Power is LOW.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.