Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26013241826013251850025

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260132418 Detector Temperature Is Low. Image quality may be affected. See LongDescription for error example Error: 260132418 Detector Temperature Is Low. Image quality may be affected. Wed Nov 9 08:45:42 2005 260132418 4 ThermalManager.cxx 1602 orp orp_subsys Wed Nov 9 08:45:41 2005 Host : Das Ermes # : 260132418 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1602 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Detector Temperature Is Low. Image quality may be affected. Current Reading:35.60 Deg C A/D Counts:176

Error: 260132518 DAS Thermal Management Has Detected That The Detector Rail Temperatures Are Out Of Tolerance For Patient/Calibration Scanning. See Long Description for Error example Error: 260132518 DAS Thermal Management Has Detected That The Detector Rail Temperatures Are Out Of Tolerance For Patient/Calibration Scanning. The Normal operating Detector temperature is 3 8 +/- 1.5 degrees C . Wed Nov 9 08:48:38 2005 260132518 4 ThermalManager.cxx 304 orp orp_subsys Wed Nov 9 08:48:37 2005 Host : Das Ermes # : 260132518 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 304 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5316832 (2 KW heater) replaces 5142966 box assy blower-heater_VCT - replaces 5124146 - used on VCT and HD See Long Descriptions for details Part #: 5316832 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Blower Heater Assembly, HD, EBM Fan, 2kW Heater FRU: Yes Status: Orderable Comments: ** FE ASSIST ** 2KW Heater assembly used on HD and newer VCT Gantries. Has a jumper plug mounted on top in front of the interface panel (see replacement procedure for illustration). The original VCT Heater assembly p/n 5119137 1KW does not have this jumper. (3/17/2009 212001261)5142966 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: box assy, blower-heater_VCT FRU: Yes Status: Orderable Comments: Per ECR 2025752, 212008898, 5/31/07. Part #: 5124146 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Box assy, blower heater-VCT FRU: Yes Status: Orderable Parts Offered as: Current: 5124146 Harvested: Not Available Exchange: Not Available Comments: Added per VCT RSPL, ms13452 041208
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues