Symptom

Scan fails aborts with PDU HVDC Buss Fault Status Scan fails aborts with PDU HVDC Buss Fault Status.

Error message

Error: 260100093 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time Mon Sep 5 13:55:52 2005 Error: 260100093 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 246 Running diagnostic: kV and mA Test: Xray Functional Test Mon Sep 5 13:55:51 2005 Host : Table Gantry Processor Ermes # : 260100093 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 246 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Static Scan: 50164/1/1 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time. Threshold 1: > 200V in 200ms Threshold 2: > 500V in 1.2s. Possible causes: HVDC Cable Short Circuit, SlipRing, or Jedi Inverter.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2218570-2 H-POWER FUSE KIT used on Lightspeed 5x systems. See Long Description for contents of fuse kit. Part #: 2218570-2 Part Type: GP CONSUMABLES Part Weight: 43.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER FUSE KIT FRU: Yes Status: Orderable Comments: Not Available Fuse KIT contents Mark Number Name Qty Description 1 2351492 2.0 200A FUSE 2 2364059 2.0 GLASS FUSE 3 46-170021P50 2.0 "FUSE 12 AMPS 250 VOLTS BUSSMAN MDA12 DUAL ELEMENT." 4 46-170021P106 2.0 FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 5 2336517-2 2.0 FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 46-170021P52 3.0 "FUSE 3 AMPS 250 VOLTS(REVIEWED TK, 7/91)." 7 46-170021P10 2.0 "006.000A 0250V 3AG FAST UL/C 8 46-170021P101 1.0 "FUSE 20 AMPS 700 VOLTS 2~ LONG X 9/16~ DIA FAST BLOW" 9 46-170021P11 2.0 "7.0A, 250V SLO-BLO FUSE. *TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 10 46-170021P31 2.0 "1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 11 2106993-5 2.0 FUSE 20 AMPS 500 VOLTS .3 SECONDS 12 46-327160P1 2.0 "12.0A, 125VAC, DUAL ELEMENT TIME DLY FNM-12, 0.406 X 1.5 LG, FIBER BODY" 13 2379651 1.0 FUSE - 700 V, 200 A, FAST SEMI
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.