Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182829

LightSpeed 3.X (Ultra, Plus, QXi) - GOC2

Symptom

Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.

Error message

Error: 182829 Remote Tilt Input to the ETC I/F Board was in an INVALID State , Remote Tilt Functionality is Disabled until ETC I/F is Reset See LD for the full error Mon Aug 8 10:35:43 2005 Error: 182829 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 651 Function: No System Function Reported Scan: 1111/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:35:44:730 Log Series: 179 Remote Tilt Input to the ETC I/F Board was in an INVALID State. Remote Tilt Functionality is Disabled until ETC I/F is Reset. Check Remote Tilt Cabling.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues