GE Healthcare · MRExcite22238172000023522217989223265120000237921822
Cannot Boot Applications Due to Host Monitor Timeout Error; reload software
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
Hostmonitor Timeout after 100 seconds 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 200002352 A HOST_SHUTDOWN was initiated by the death of ifcc 2217989 Internal Software Error Unable to initialize the Raw Data Base Manager Tue Dec 27 12:00:51 2005 Error: 2217989 Host: mroc Process: ifcc File: ifccMain.cc Line: 611 Internal Software Error. Unable to initialize the Raw Data Base Manager. Status = 2232651 200002379 Process ifcc pid 21822 exited with status=1 Tue Dec 27 12:00:51 2005 Error: 200002379 Host: mroc Process: hostMonitor File: ssSigHandlers.c Line: 136 Process ifcc, pid 21822 exited with status=1 2224330 tn_init error, errno = 2 No such file or directory Error: 2224330 Host: WHC_MR1 Process: NSP File: TardisReset.cc Line: 487 tn_init error, errno = 2 No such file or directory 200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. s
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.