Pular para o conteúdo principal
GE Healthcare · MRSigna2241523

Will not scan with Gradient Overload error 2241523 Due to GPM

Symptom

Will Not Scan with Gradient Errors.

Error message

2241523 The GIP Reported a Long Overload Fault on the X Axis of the GPM. Tue Apr 20 20:24:33 2004 Error: 2241523 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2324 The GIP reported a Long Overload Fault on the X Axis of the GPM. Possible causes: X Axis GPM. Try swapping the GPM Axis connections to isolate the problem.->

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2151183 GRADIENT POWER MODULE For SGD. 8651 Is SGD Hi-Slew Gradient Power Module.
Area
MR

Full cause & step-by-step resolution