Pular para o conteúdo principal
GE Healthcare · CT00000072

ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

XG Error: 117 00000072 Modification Parameter is not allowed XG117 00000072 Modification Parameter is not allowed

Applies to

Product
ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) ProSpeed Qj and Zj (Lemage, Qj, Qj-Base, Qj-Hlt) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
P# 2119708-2 Filament Control Board for YMS Prospeed Systems (old part # 2119708) FE should check/replace the C17 on filament board, 470uf 63volts before replacement of this board - also check C40! This is the primary failure part on this board. C17 has been replaced with 470uf 80 volts purchased locally and this fixed the filament errors. Also check C40. Part #: 2119708-2 Part Type: GP CONSUMABLES Part Weight: 0.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILAMENT HEATER BOARD FRU: Yes Status: Orderable Parts Offered as: Current: 2119708-2 Harvested: Not Available Comments: FLS - FILAMENT CONTROL BRD U/O CT QJ & CT MINI REPLACES 2119708 PER PCN P 10/99 02 ON CONSUMABLE COLLECTION PER G TOURNE 1/16/04 MS5277 DO THE CHANGE UP ONCE OLD MATERIAL PULLED BACK DO NOT DISCOUNT - PER SHAKUN CHECK CAPACITORS: C17 AND C70 GELS - 17-09-2003 20PCES WILL SEND TO BUC FOR TEST AND 30 PCES IN FW 39 FLORIANE 16/09/03,PCB CONT FIL
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues