Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017801230023045230017823038304900000

LightSpeed 5.X (60kW)

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017823 Error: Invalid ratio range detected during the Fast Cal See the long description for the full error Thu Aug 26 02:56:40 2004 Error: 230017823 Host: ct01 Process: accs File: collSweepProcessing.c Line: 2436 ACCS CCal: Error: Invalid ratio range detected during the Fast Cal. The minimum ratio is 1.038304 and is great than the maximum ratio 0.900000

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required