GE Healthcare · Nuclear MedicineInfinia1040111401
Infinia II Unable to open some protocols (for scan or with Protocol Editor) but able to open some others.
Symptom
Infinia II Unable to open some protocols (for scan or with Protocol Editor) but able to open some others. Even Factory Wholebody Bone Scans cannot be opened. Infinia II Unable to open some protocols (for scan or with Protocol Editor) but able to open some others. Even Factory Wholebody Bone Scans can not be opened. Error message seen 'failed to open protocol error 10401, load sequence error, use default or restore'.
Error message
Infinia System error message" The value of the gantry bed height position is illegal...." during a simple Wolebody scan Infinia System error message" The value of the gantry bed height position is illegal...." during a simple WB scan. ERROR_PROTOCOL_ACTION_FAILED, 10401, ERROR Failed to open protocol, load sequence error Example Error: Messages Chain: DEVELOPMENT_ERROR_MESSAGE=Failed to perform load operation on protocol EVOLUTION ONE DAY USER_ERROR_MESSAGE=Failed to open protocol USER_FIX=1) Retry the last operation. 2) In the protocol editor, delete the protocol and recreate it. 3) If this is a factory protocol, restore the protocols from your latest backup. CODE=(ERROR_PROTOCOL_ACTION_FAILED,10401,ERROR) SOURCE=EVOLUTION ONE DAY TIME_CREATED=Fri Dec 19 07:29:40 CST 2008 Caused by: DEVELOPMENT_ERROR_MESSAGE=Failed to load sequence USER_ERROR_MESSAGE=Failed to load sequence USER_FIX=Use the Restore utility to restore this file CODE=(ERROR_SEQUENCE_LOADING,11401,ERROR) SOURCE=com.ge.med.fi.tacq.repository.SysDataAdapter@4eff2c
Applies to
- Models
- Infinia
- Product
- Infinia II Infinia II Acquisition System
- Software
- Infinia II 2.005.005.5 (2.55) Infinia II 2.005.005.5 (2.55) Infinia II 2.005.005.5 (2.55)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.