GE Healthcare · CTLightSpeed23000006929144370005000324400036000
Scanning aborted WITH TAXI data violation errors
Symptom
"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors
Error message
Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical Scan: 29144/9/3 Scan Seq Id: 37000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 10 1 View Data Corruption Detected: ---------------------- Bad View Details: Bit Count Event description 0 1 Checksum Field Error 1 1 View Number Field Error 2 1 Row Number Field Error 3 1 Data Length Field Error 4 1 Data Type Field Error 5 1 Magic Number Field Error 7 1 Consecutive Corrupted Segments Detected Mon Dec 22 07:16:59 2008 Error:
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2333615 850MB TRANSMITTER used on LightSpeed 16 slice Systems 2333615 850MB TRANSMITTER used on LightSpeed 4.X (16-slice) Systems 2333615 850MB TRANSMITTER used on LightSpeed 4.X (16-slice) Systems Part #: 2333615 Part Type: GP CONSUMABLES Part Weight: 0.46 LBR Repairable: No Returnable: No Hazardous Material: N Description: 850 MB TRANSMITTER FRU: Yes Status: Orderable Comments: FLS - 850MB TRANSMITTER ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required