Pular para o conteúdo principal
GE Healthcare · CTHiSpeed186403186507

CT/e Dual, CT ProSpeed AII/FII Irix or Linux

Symptom

Hardware stop scan, collimator error, Aperture width is out of specification See Long D for details Hardware stop scan, collimator error, Aperture width is out of specification Often this happens just after servicing the Collimator or replacing an X-ray tube. But not always.

Error message

Error: 186403 OGP did not receive READY status from CIF for z-axis collimator control or OGP detected Aperture is not ready See LD for complete error Mon May 9 18:03:28 2005 Error: 186403 Host: CTE Process: tgp_out File: tgp_errors.c Line: 317 Aperture Error. Code : 20-1020-04 Error detected by : OGP OGP did not receive READY status from CIF for z-axis collimator control or OGP detected Aperture is not ready.

Error: 186507 Z-Axis Collimator Error. Code : 30-1111-07 The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning. See LD for complete error Mon May 9 18:03:19 2005 Error: 186507 Host: CTE Process: tgp_out File: tgp_in_input.c Line: 449 Z-Axis Collimator Error. Code : 30-1111-07 Error detected by : CIF The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning.

Applies to

Models
HiSpeed
Product
CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2201179 Collimator Z-axis motor driver board See LD for more info 2201179 Collimator Z-axis motor driver board Part #: 2201179 Part Type: GP CONSUMABLES Part Weight: 2.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9233SC MOTOR DRIVER FRU: Yes Status: Orderable Comments: FLS - MOTOR DRIVER U/O CT HISPEED-YMS GANTRY M3 FW20 '98 MS1101 3/98
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.