Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1108039565112831100911046

Site cannot send a specific exam or series to PACS or other dicom hosts.

Symptom

Site cannot send a specific exam or series to PACS or other dicom hosts. All other exams send without error. See Long Description for more details Site cannot send a specific patient to PACS or other dicom host. All other exams send without error. Error as seen in the GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11046 Host: fchct1 Process: dcserver File: dcspush.c Line: 1381 STORE response received with refused status (0xA700) - Aborting! Fri Apr 7 19:00:01 2006 Error: 11009 Host: fchct1 Process: dcserver File: dcspush.c Line: 1516 Association was aborted with PACS_Live Fri Apr 7 19:00:01 2006 Error: 11080 Host: fchct1 Process: dcserver File: dcscommon.c Line: 242 Connection error: 11283 (PACS_Live) Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1381) [Dcserver]> STORE response received with refused status (0xA700) - Aborting! MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1516) [Dcserver]> Association was aborted with PACS_Live MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcscommon.c:242) Connection error: 11283 (PACS_Live)

Error message

Error: 11080 Connection Error See the Long Description for the full error code as seen in gesyslog Thu Dec 29 23:00:01 2005 Error: 11080 Host: wct1 Process: dcserver File: dcscommon.c Line: 242 Connection error: 39565 (EFILM-WHRA) as seen in DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcscommon.c:242) Connection error: 11283 (PACS_Live)

Error: 11009 Association was aborted with hostname See LD for full details Error: 11009 Association was aborted with hostname Error as seen in GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11009 Host: fchct1 Process: dcserver File: dcspush.c Line: 1516 Association was aborted with PACS_Live Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1516) [Dcserver]> Association was aborted with PACS_Live

Error: 11046 STORE response received with refused status (0xA700) - Aborting! See LD for full text. Error: 11046 STORE response received with refused status (0xA700) - Aborting! Error as seen in the GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11046 Host: fchct1 Process: dcserver File: dcspush.c Line: 1381 STORE response received with refused status (0xA700) - Aborting! Error as seen in the DCSLOG MESSAGE from 

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required