Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260001008260004025502291912726000104950133

Collimator/DAS reset failed

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB (could be the CCB) sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS

Error: 260004025 DAS / Collimator firmware object: Lost communications with node applications. See Long Description for full error Thu Jan 13 06:58:23 2005 Error: 260004025 Host: OBC Process: Collimator File: col_cf_in_mail.c Line: 442 Function: No System Function Reported Scan: 50229/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:58:23:235 Log Series: 47 Collimator firmware object: Lost communications with node applications. Thu Dec 31 11:45:03 2009 Error:260004025 Host: OBC Process: DAS File: das cf in mail.c Line: 430 Function: No System Function Reported Scan: 19127/2/0 Type: None/Unknown Exception Level: Pri/

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution