Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2687225588

LightSpeed 3.X (Ultra, Plus, QX/i) -Octane

Symptom

System posts: Edit Patient Data cannot remove the exam entry: it may be in use. System posts Edit Patient Error message: Edit Patient Data cannot remove the exam entry: it may be in use.

Error message

Error: 1031 stored procedure failed See Long Description for Details. Sun Nov 14 20:00:37 2004 Error: 1031 Host: mcas Process: dbrserver stored procedure failed : -284 - getexamuidinfo failed Also check usr/g/ctuser/logfiles/epdlog with command: tail -100 /usr/g/ctuser/logfiles/epdlog | more MESSAGE from Process 26872>> Sun Nov 14 20:12:15 2004 (netclient.c:93) [Client]> The request manager started MESSAGE from Process 26872>> Sun Nov 14 20:12:48 2004 (:0) [EPD] Failed to remove the exam entry Thu Jul 17 13:45:18 2008 Error: 1031 Host: ct99 Process: dbrserver File: batch.ec Line: 599 stored procedure failed : -243 - Fetch failed for image - ISAM code : -143 Wed Jul 2 10:54:38 2008 Error: 1031 Host: CT01 Process: dbwserver File: batch.ec Line: 934 stored procedure failed : -25588 - Identifyimage failed - ISAM code : 22, object , series $

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.