Pular para o conteúdo principal
GE Healthcare · MRSignaHiSpeedExcite222593522253342225939222533722275562225944

PDU in Stop Mode and TPS Download failure 2224733 2225334 .

Symptom

PDU in Stop Mode and TPS Download failure

Error message

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

2225334 SRI failed to respond to a data packet from SPI

2225939 The PAC did not respond correctly to a reset command from IPG. Error: 2225939 Host: rmr2 Process: NSP File: HPC_PORT_ID:spu_os_init Line: 469 The PAC did not respond correctly to a reset command from IPG. Check PAC power and fiberoptic connections, run PAC diags.

2225337 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Error: 2225337 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri watchdog fail Line: 13 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Watchdog packets waiting for a response: 0x0001

2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. ***

2225944 MRconfig.cfg and PAC HW 

Applies to

Models
Signa, HiSpeed, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only