GE Healthcare · CTLightSpeed24521651000200002352
LightSpeed 4.X - GRE
Symptom
Scanner shuts down on its own during scanning. Scanner shuts down on its own during scanning
Error message
Error: 245216 Scanmgr was unable to deallocate a pre-allocated but unusuable scanfile because it was locked by other processes (probably recon). See LD for complete error Wed Apr 5 12:29:51 2006 Error: 245216 Host: gect_16 Process: scanmgr File: start_finish_request.cxx Line: 244 Function: System Control : Scan Control Scan Type: Axial Scan: 980/200/5 Scan Seq Id: 51000 Scanmgr was unable to deallocate a pre-allocated but unusuable scanfile because it was locked by other processes (probably recon). It will be deallocated by the last process who unlocks it. Error: 200002352 A HOST_SHUTDOWN was initiatied by the death of examRxEnv. See LD for complete error Wed Apr 5 12:02:45 2006 Error: 200002352 Host: gect_16 Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiatied by the death of examRxEnv.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.