GE Healthcare · CTLightSpeed26000650050000
LightSpeed 4.X - GRE
Symptom
Unable to scan or perform tube warmup When the tube rotor spins up or down it may sound like a jet plane is landing and the gantry covers vibrate
Error message
Error: 260006500 The Axial AC Drive reported fault code : F2- Auxiliary Fault The interlock between the Chopper control Circuit on the axial I/F Board and the Drive is Open See LD for complete error Wed Mar 8 07:57:35 2006 Host: STC Proc: Axial Error: 260006500 File: Ax2Fault.c Line: 196 Function: Axial Control : Axial Control Scan: 50000/1/0 Type: None / Unknown Exception Level : Pri / Most Servere Time: 07:57:35:910 Log Series: 45 The Axial AC Drive reported fault code : F2-Auxiliary Fault The interlock between the Chopper control Circuit on the axial I/F Board and the Drive is Open
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P104 : 10 Amp 700vdc Fuse - used in HEMRC and Axial Dynamic Brake and others See Long Description for more info 46-170021P104 10 Amp 700vdc F1 Fuse -used in HEMRC and Axial Dynamic Brake and others many pictures below Part #: 46-170021P104 Part Type: GP CONSUMABLES Part Weight: 11.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 10 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 10 AMPS 700 VOLTS ULTRA FAST. USED ON FAST RHAPSODE 7/14/97 GWENN 08992 Fuse for Axial Brake Module Fuse for HEMRC Rotor controler (HSA-1.x) Fuse for HEMRC Rotor Control (2.x-4.x)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.