GE Healthcare · Nuclear MedicineInfinia
VG gantry rotation direction is unreliable, when commanded to go clockwise randomly goes other direction with no error i…
VG gantry rotation direction is unreliable, when commanded to go clockwise randomly goes other direction with no error in panel
Symptom
VG gantry rotation direction is unreliable, when commanded to go clockwise randomly goes other direction with no error in panel
Applies to
- Models
- Infinia
- Product
- VGH VG and VG Hawkey and Varicam
- Component
- *Gantry Gantry Gantry B010200028 VG rotation motor B010200028 VG rotation motor Part is obsolete see ITF 5408614-1EN for ordering part UGP001489 - Infinia SAX driver. 5409135 - VG rotation motor SP, this is the Infinia motor but with longer cables, 2 m instead of 1 m. 5408499 - set of brackets and transformer required for replacement of Bru 200 by SAX. Upon releasing of the parts Haifa will release also a Service Note.
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.