Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed230023045230017801230017823038304900000

Fastcals failed to complete

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230017823 Error: Invalid ratio range detected during the Fast Cal See the long description for the full error Thu Aug 26 02:56:40 2004 Error: 230017823 Host: ct01 Process: accs File: collSweepProcessing.c Line: 2436 ACCS CCal: Error: Invalid ratio range detected during the Fast Cal. The minimum ratio is 1.038304 and is great than the maximum ratio 0.900000

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.