GE Healthcare · MRSigna22256932225702
Will Not Scan with 2225693 and 2225702 MDS Errors
Symptom
Will Not Scan with Gradient Errors.
Error message
2225693 MDS link failure. An MDS device failed to respond to IPG. Wed Oct 29 00:14:08 1997 2225693 1 CMR1 OC0 HPC_PORT_ID PMH.m 1.29 spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0252 Transmitted header: 0x10 Transmitted opcode: 0x02 2225702 MDS link failure. The link appears to be broken between the IPG and the Gradient Subsystem. Fri Feb 18 05:47:41 2005 Error: 2225702 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_grad Line: 100 MDS link failure. The link appears to be broken between the IPG and the Gradient Subsystem. Check the connections and signal level of the MDS fiber optic link between the TYME II card and the Gradient Subsystem. Run MDS fiber optic link diagnostics.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.