GE Healthcare · CTLightSpeed1831212257550013
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183121 FIL/MA fault detected: Open filament circuit. See the long descripton for the full error code Tue Oct 11 13:12:27 2005 Error: 183121 Host: OBC Process: Generator Function: XRAY GENERATION : Filament Selection Scan: 22575/2/1 Type: Helical Exception Level: Pri/Most Severe FIL/MA fault detected: Open filament circuit. Desired focal spot = 0. (0 = large, 1 = small) <== Large Filament is open OR Fri Sep 22 07:03:11 2006 Error: 183121 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 341 Function: XRAY GENERATION : Filament Selection Scan: 50013/10/40 Type: Axial Exception Level: Pri/Most Severe Time: 07:03:11:045 Log Series: 393 FIL/MA fault detected: Open filament circuit. Desired focal spot = 1. (0 = large, 1 = small)<-------- Small
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-136370P23 K78 FILAMENT RELAY See long description for details Part #: 46-136370P23 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 024VDC 2P2T 028VDC 10AMP FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.