LightSpeed 5.X (60kW)
Symptom
After Fast Cal completes software begins a Z-Slope Calibration Test. See long description for details After Fast Cal completes software begins a Z-Slope Calibration Test. Screen states Generating Cal vectors, May Take 4 Minutes but the test does not complete.
Error message
Error: 230023041 Found the Zslope cal vectors are already 6 month old and need to update See the Long Description for the full error code Mon Oct 17 23:58:54 2005 Error: 230023041 Host: cts1 Process: dailyPrepRx File: DailyPrepUtils.c Line: 4394 Found the Zslope cal vectors are already 6 month old and need to update Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. See LD for Full Error Error: 230017124 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000. Tue Dec 27 14:51:03 2005 Error: 230017124 Host: ct01 Process: accs File: zslope.c Line: 1910 Read minimum over maximum ratio from ZslopeCal.cfg is 0.550000.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.