Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260107204

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

The positioning lights will not come on when the Patient was being positioned for a scan The laser lights will not come on when the Patient was being positioned for a scan

Error message

Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Description for full error Wed Apr 7 21:18:13 2004 Error: 260107204 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 230 Wed Apr 7 21:18:13 2004 Host : Table Gantry Processor Ermes # : 260107204 Exception Class : Abort Severity : Secondary File : amc_mgr.c Line# : 230 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
46-170021P11 7.0A 250V SLO-BLO FUSE used in Hpower Axail Dynamic Brake Module 2357524 46-170021P11 7.0A 250V SLO-BLO FUSE used in Hpower Axail Dynamic Brake Module 2357524 46-170021P11 7.0A 250V SLO-BLO FUSE used in Hpower Axail Dynamic Brake Module 2357524. see attachment 1 and 2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.